Application
1.Installation of lighting tube, outside the studio can clearly see the situation; needling / extrusion speed 10 ~ 80mm / s can be adjusted; - Needling / extrusion force value
250N ~ 13KN can be set to meet the requirements of different standards
2.Needling/Extruding force value 250N~13KN can be set arbitrarily to meet different standard requirements.
3.Wire coat metal fire pipe, effectively prevent the battery test process of ignition and combustion phenomena;
4.separation of control and test box with form, available at 1~2 meters operation, be safe.
5.Squeeze: The test cell is squeezed in two planes, and a squeezing pressure of about 13KN is applied through a vise or a hydraulic arm with a piston diameter of 32mm, the squeezing continues until it rises, and once the maximum pressure is reached, the squeezing is lifted.
6.Needling: The test shall be carried out at an ambient temperature of 20℃±5℃, the battery connected with a thermocouple (thermocouple contacts are fixed on the surface of the battery) is placed in a fume cupboard, and a non-corrosive rust-free steel needle with a diameter of 2-8mm is used to pierce the center of the largest surface of the battery at a speed of 10mm/s-40mm/s and held for an arbitrary period of time. center of the largest surface of the cell at a speed of 10mm/s-40mm/s and hold it for an arbitrary period of time.
Squeeze Indicator
controllers | 7-inch touch screen |
Test area space | 250mm wide x 300mm deep |
Outer box size | Approx. 750*750*1800mm (W*D*H) subject to actual size |
drive method | motor drive |
Strength range | 1~20kN (adjustable) |
Force Measurement Accuracy | 0.1% |
unit conversion | kg, N ,lb |
Squeeze Stroke | 300mm |
Force value display | PLC touch screen display |
Battery Squeeze Head | Standard extrusion head, area ≥ 20cm². |
Inner box material | Stainless steel 1.5mm thick |
Outer case material | 1.2 mm thick A3 cold plate with lacquered finish |
safety device | The back side of the box is designed with air vent and pressure relief device 250*200mm, the box is equipped with lighting |
viewing window | 250x200mm two-layer vacuum toughened glass viewing window with explosion-proof grill |
exhaust vent | The back of the box is equipped with a high temperature exhaust fan and reserved exhaust pipe interface φ150mm.
The exhaust fan will turn on and work immediately after the power is turned on. |
box door | Single door, left opening |
Box door switch | Open-on-disconnect threshold switch ensures no misuse and personnel safety. |
caster | Four universal castors underneath the machine for free movement. |
Acupuncture Indicator
steel needle | Φ3mm/φ5mm high-temperature resistant tungsten steel needle, length 100mm (can be specified) 2pcs each. |
Needle Stroke | 200mm |
unit conversion | Kg、N 、lb |
Needling speed | 10 ~40mm/s (adjustable) |
Needlepoint force value | 1~300Kg |
Force value display | PLC touch screen display |
drive method | Motor control, adjustable speed |
Data Acquisition
Voltage Acquisition |
Voltage range: 0~100V |
Acquisition rate: 200ms |
|
Acquisition channel: 1 channel |
|
Accuracy: ±0.2%F.S (0~100V) |
|
temperature acquisition |
Temperature range: 0℃~1000℃ K-type thermocouple |
Acquisition rate: 200ms |
|
Acquisition channel: 1 channel |
|
Accuracy: ±2℃ |